MedWOW / أجهزة طبية / التصوير / جهاز تصوير بالرنين المغنطيسي / Siemens / MAGNETOM Open Viva / مواصفات المصنع
مواصفات المصنع - MAGNETOM Open Viva, Siemens
لاحظ من فضلك: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
| MODEL | MAGNETOM Open Viva |
| MAGNET | ResistiveC-shaped |
| POWER NEEDED | |
| Line voltage, V (POWER NEEDED) | 380/400/420/440/480 |
| Kva (POWER NEEDED) | 35 |
| CLINICAL USE (POWER NEEDED) | Whole body |
| CARDIAC GATING | |
| Ecg/peripheral (CARDIAC GATING) | Yes/yes |
| IMAGING MODES | |
| Single (EYEPIECE) | Yes |
| Multislice (IMAGING MODES) | Yes |
| Volume study (IMAGING MODES) | Yes |
| SURFACE COILS | |
| Spine (SURFACE COILS) | CP flexible spine |
| Knee (SURFACE COILS) | 21cm multipurpose |
| Neck (SURFACE COILS) | 16/21 cm multipurp |
| Tmj (SYSTEM CAPABILITIES) | |
| Extremity (SURFACE COILS) | 16/21 cm multipurp |
| Head (SURFACE COILS) | CP head |
| Shoulder (SURFACE COILS) | 16/21 cm multipurp |
| Additional (Alarms, high/low) | 45cm multipurpose, 35cm multipurpose for MR-guided procedures |
| BORE DIAMETER or W x H, cm (SURFACE COILS) | MRT (Open Magnet) |
| DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC) | Optional |
| GRADIENT SUBSYSTEM | |
| Strength, mt/m (GRADIENT SUBSYSTEM) | 15 |
| Rise time to 10mT/m, msec (GRADIENT SUBSYSTEM) | 0.6 |
| Cpu (NETWORK/DATABASE) | Turbo SPARC-based host |
| Memory size, mb (GRADIENT SUBSYSTEM) | 64 |
| ARRAY PROCESSOR (EXTERNAL BEAM) | SMI-5 (Siemens Medical Imager) |
| Memory size, mb (GRADIENT SUBSYSTEM) | 146 |
| Storage media/size (GRADIENT SUBSYSTEM) | 1.7GB MOD drive |
| Image storage capacity (GRADIENT SUBSYSTEM) | ~11,000 (256 x 256) |
| ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD) | 0.2 |
| OPTIONS | PACSnet Ethernet, MR-guided procedures kit, comfort kit, cardiac evualuation package, extra memory, hard disk expansion, Prominence work- station |
| SHIMMING (CRYOGEN USE, L/hr) | Passive, active |
| Diameter, cm (VACUUM) | |
| Body coil (DIAMETER, cm) | 41pole to pole |
| Head coil (DIAMETER, cm) | 20.3x 17 |
| Reconstruction time | |
| Single slice, sec (RECONSTRUCTION TIME) | <0.65 |
| Multislice, sec (RECONSTRUCTION TIME) | <0.65/slice |
| Volume, sec (RECONSTRUCTION TIME) | <0.65/slice |
| Cooling method (X-RAY TUBE) | Closed loop, chilled water |
| Chilled h2o, l/min (POWER NEEDED) | 17-40, varies w/temp |
| MAGNET WEIGHT, kg (CRYOGEN USE, L/hr) | 12100 |
| DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr) | 195x 120 x 200 |
| Angiography | Optional MT, TONE, 2-D & 3-D TOF; Turbo MRA, PC (WIP) |
| IMAGING | |
| Pulse sequences (IMAGING) | SE, IR, True IR, Turbo IR, GE, 2-D & 3-D FLASH, 2-D & 3-D FISP, 3-D PSIF, 3-D DESS, 2-D True FISP, TSE, Turbo FLASH, combination echo |
| Repetition time, msec (IMAGING) | 28minimum SE, 11minimum GE |
| Echo time, msec (IMAGING) | 15mininum SE, 4minimum GE |
| Inversion time, msec (IMAGING) | 35minimum |
| Slice thickness, mm (GANTRY) | 2-60 (2-D), 0.5-50 (3-D) |
| Fov, cm (IMAGING) | 4-40 |
| Fov offsets (IMAGING) | ±20 |
| Scan orientations (IMAGING) | Axial, sagittal, and coronal orthogonal; oblique; double oblique; MAO |
| Measuring matrix (IMAGING) | Up to N x 1024, N = 32-512 in increments of 1 |
| Display matrix (IMAGE DISPLAY) | 1024x 1024 |
| Pixel intensity (IMAGING) | 1-4,096 |
| Spatial resolution, mm (IMAGING) | 0.31,in plane |
| BORE FEATURES (DIAMETER, cm) | Open 270 deg patient access, removable table, intercom, nurse-call bulb, opt music headphones |
| CONFIGURATION | Open |
| OTHER ATTRIBUTES (Interference compensation) | Integrated MIP and MPR; standard integrated image noise filters, VCR interface, cine display, remote diagnostics package, quantitative analy- sis; 2 coils plug in simultaneously. |
| FDA CLEARANCE (Interference compensation) | Yes |
| CE MARK (MDD) (Interference compensation) | Yes |
| MARKETING REGION (Interference compensation) | Worldwide |