مواصفات المصنع - AIRIS II, Hitachi
لاحظ من فضلك: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
| MODEL | AIRIS II |
| MAGNET | Permanent |
| POWER NEEDED | |
| LINE VOLTAGE, VAC | 190-240, single phase |
| Kva (POWER NEEDED) | 3 |
| A/c, btu/hr (POWER NEEDED) | <10,000 |
| CLINICAL USE (POWER NEEDED) | Whole body |
| OPTIONS | DualQUAD multiple array package, UPS, VCR interface and joint motion packages, remote monitor, laser imager, iV/iO package |
| IMAGING MODES | |
| Single (EYEPIECE) | Yes |
| Multislice (IMAGING MODES) | Yes |
| Volume study (IMAGING MODES) | Yes |
| Additional (Alarms, high/low) | Multislab, multiangle, dual slice, gating |
| Reconstruction time | |
| Single slice, sec (RECONSTRUCTION TIME) | 0.05 |
| Multislice, sec (RECONSTRUCTION TIME) | 0.05/slice |
| Volume, sec (RECONSTRUCTION TIME) | 0.05/slice |
| SURFACE COILS | |
| Spine (SURFACE COILS) | MA CTL |
| Knee (SURFACE COILS) | Quad knee |
| Neck (SURFACE COILS) | QD C-spine |
| Tmj (SYSTEM CAPABILITIES) | Bilat solenoid |
| Extremity (SURFACE COILS) | Latchable extremity |
| Head (SURFACE COILS) | QD head |
| Breast (APPLICATORS) | Bilateral QD breast |
| Shoulder (SURFACE COILS) | MA shoulder |
| Additional (Alarms, high/low) | QD flex body (S,M,L, XL), MA head/neck, QD wrist, QD latchable coils (3 sizes) |
| BORE SIZE, dia x L | |
| Or w x h x l, cm (BORE SIZE, dia x L) | 110 x 43, open |
| ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD) | 0.3 |
| SHIMMING (CRYOGEN USE, L/hr) | Passive & 3-axis per patient auto shim |
| Diameter, cm (VACUUM) | |
| Body coil (DIAMETER, cm) | 33, 38, 48, 61; flexible quadrature |
| Head coil (DIAMETER, cm) | 29x22 split-top quad |
| Respiratory gating (DISPLAY) | Yes |
| GRADIENT SUBSYSTEM | |
| Strength, mt/m (GRADIENT SUBSYSTEM) | 15 |
| Rise time to 10 mt/m, msec (GRADIENT SUBSYSTEM) | 0.6 |
| Cpu (NETWORK/DATABASE) | 64-bit RISC based |
| Memory size, mb (GRADIENT SUBSYSTEM) | 1 GB |
| ARRAY PROCESSOR (EXTERNAL BEAM) | Hitachi |
| Memory size, mb (GRADIENT SUBSYSTEM) | 256, optional 512 |
| Storage media/size (GRADIENT SUBSYSTEM) | Magnetic disk, optical disk |
| Image storage capacity (GRADIENT SUBSYSTEM) | 27,000 magnetic disk, 17,000 optical disk |
| IMAGING | |
| Pulse sequences (IMAGING) | SE, GE, GR, IR, FIR, STIR, FLAIR, FSE, DE-FSE, SS-FSE, SARGE, BA-SARGE, RSSG, TRSG, TOF MRA, MTC, MR fluoro, msEPI, DWI, FatSeparation S & G |
| Repetition time, msec (IMAGING) | 7-16, 700 |
| Echo time, msec (IMAGING) | 3.5-2,000 |
| Inversion time, msec (IMAGING) | 8-00 |
| Slice thickness, mm (GANTRY) | 2-100 (2-D), 0.5-5 (3-D) |
| Fov, cm (IMAGING) | 42 |
| Fov offsets (IMAGING) | Up to maximum FOV |
| Scan orientations (IMAGING) | Transverse, coronal, sagittal, single oblique, multiple oblique angles |
| Measuring matrix (IMAGING) | 64-1,024 steps of 1, freq x 64-512 (steps of 4-phase), 1/2 & 3/4 scan rect |
| Display matrix (IMAGE DISPLAY) | 1280 x 1024 |
| Pixel intensity (IMAGING) | -2,000 to +4,000 |
| Resolution, mm (PERFORMANCE) | 0.2 |
| DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC) | Storage, query, retrieve, print |
| CARDIAC GATING | |
| Ecg/peripheral (CARDIAC GATING) | Yes/yes |
| Angiography | 2-D/3-D TOF, CE-MRA |
| BORE FEATURES (DIAMETER, cm) | Open bore, short depth, intercom, SoftSound gradients, technologist alert |
| MAGNET WEIGHT, kg (CRYOGEN USE, L/hr) | 15700 |
| DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr) | 194 x 273 x 179 |
| ECHO PLANAR IMAGING (CARDIAC GATING) | Multishot |
| 5-GAUSS FRINGE FIELD | |
| Radial/axial, m (5-GAUSS FRINGE FIELD) | <2.5/2.2 |
| CONFIGURATION | Open |
| OTHER ATTRIBUTES (Interference compensation) | Low-conductance pole piece with minimal eddy currents; digital RF system quadrature transmitter and up to 4 receiver preamplifiers; automatic and user-variable receiver bandwidth; cine display; modular CPU architecture with remote service; real-time multi |
| FDA CLEARANCE (Interference compensation) | Yes |
| CE MARK (MDD) (Interference compensation) | Yes |
| MARKETING REGION (Interference compensation) | Asia, Europe, North America |