X-RAY TUBE ANODE | |
Heat storage, hu (X-ray tube anode) | 8,0,000 |
Heat dissipation rate, hu/min (X-RAY TUBE) | 1782000 |
Tube focal spot, mm (X-ray tube anode) | 0.7 x 0.6, 0.9 x 0.9 |
POWER NEEDED | 460/480 VAC nominal, 50/60 Hz, 3-phase delta or Wye |
Type (CONFIGURATION) | Multislice |
# of slices (X-ray tube anode) | 64 |
Image storage | |
Hd capacity, gb (IMAGE STORAGE) | 146 |
FLOOR SPACE NEEDED | |
Size, m2 (FLOOR SPACE NEEDED) | 28 minimum |
SCORING OF CORONARY ARTERY CALCIFICATION (IMAGE STORAGE) | Optional |
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC) | Yes |
DISPLAY | |
MONITOR | 20" (2), opt flat |
Matrixes, pixels (DISPLAY) | 1280 x 1024 |
Range of ct numbers (DISPLAY) | -1,024 to +3,071 |
Image enlarging scale (DISPLAY) | Up to 8x |
Max # slices displayed simultaneously (DISPLAY) | 16 |
GENERATOR | |
Output, kw (X-RAY GENERATORS) | 100 |
Ma range (IMAGING SYSTEM) | 10 |
TABLE FEATURES | |
Vertical, cm (Range of movement) | 51-107 |
Max. patient weight, (precision), kg (Range of movement) | 180 (±0.25 mm), 205 (±1 mm) |
HELICAL SCANNING (DISPLAY) | Yes |
Max scan time, sec (DISPLAY) | 120 |
Max scan volume, cm (DISPLAY) | 170 |
Spatial resolution, lp/cm (DISPLAY) | Same as axial |
Pitch (DISPLAY) | 0.5625:1, 0.9375:1, 1.375:1, 1.75:1 |
Reconstruction time per image, sec (DISPLAY) | 0.063 |
PERFORMANCE | |
Minimum interscan time, sec (PERFORMANCE) | 1 |
Dynamic scan pace (PERFORMANCE) | 960 scans/min |
High-contrast spatial resolution 0% mtf, lp/cm (PERFORMANCE) | 15.4, 19.6-Z |
Low-contrast resolution, mm at % at <=4 rads (PERFORMANCE) | 5 at 0.3% at 1.3 mGy 8" CATPHAN |
Sound, % at <= 2.5 rads (PERFORMANCE) | 0.32 at 2.85 rads |
Image reconstruction matrixes (IMAGE PROCESSING) | 512 x 512 |
Reconstruction time | |
For localization scan, sec (Reconstruction time) | Real time |
GANTRY | |
Geometry (GANTRY) | Rotate-rotate, slip ring |
DETECTOR (SCATTERED LIGHT) | HiLight ceramic |
Rows (GANTRY) | 64 |
Elements/row (GANTRY) | 912 |
# detection channels (GANTRY) | 64 x 912 |
Rotation times, sec 360 (GANTRY) | 0.4,0.5,0.6,0.7,0.8,0.9,1,2,3, 4; optional 0.35 |
Slice thickness, mm (GANTRY) | 0.63,1.25,2.5,3.75,5,7.5, 10 |
X-ray fan beam angle, ° (GANTRY) | 56 |
Gantry angle deg (GANTRY) | ±30 |
Gantry size, hxwxd,cm (GANTRY) | 188.2 x 222.5 x 100.6 |
FDA CLEARANCE (Interference compensation) | Yes |
CE MARK (MDD) (Interference compensation) | Yes |
MODEL | LightSpeed VCT |
High-contrast spatial resolution | 0.35 mm |
0% mtf, lp/cm (High-contrast spatial resolution) | 15.4, 19.6-Z |
50% mtf, lp/cm (PERFORMANCE) | 10.2, 19.6-Z |
Slice sensitivity profile | |
Low-contrast resolution, mm at % at ≤20 mGy (2 rads) (Slice sensitivity profile) | 5 mm at 0.3% at 1.3 mGy; 8“ CATPHAN |
Noise, % at ≤25 mGy (2.5 rads) (Slice sensitivity profile) | 0.32% at 2.85 rads |
Range of movement | |
Longitudinal, cm (Range of movement) | 170, optional 200 |
Scanning range,cm (Range of movement) | 170, optional 200 |
Max load capacity (accuracy, mm), without restrictions, kg (Range of movement) | 227 (±0.25 mm) |
Gantry weight, kg (GANTRY) | 1906 |
Gantry opening, cm (GANTRY) | 70 |
Scan localizer (GANTRY) | Laser |
X-RAY TUBE | |
Heat storage, MHU (X-RAY TUBE) | 8 |
Heat dissipation rate, kHU/min (X-RAY TUBE) | 1782 |
Tube cooling (X-ray tube anode) | Oil/air |
Tube focal spots, mm (X-RAY TUBE) | 0.7 x 0.6, 0.9 x 0.9 |
Expected tube life, scan sec (& calendar) (X-RAY TUBE) | See tube warranty |
Maximum ma (X-ray tube anode) | 800 |
Maximum scan time at maximum ma, sec (X-ray tube anode) | 5 |
IMAGE PROCESSING (MAXIMUM DISPLAY) | |
Standard or optional (IMAGE PROCESSING) | Optional |
Recommended post-processing workstation (IMAGE PROCESSING) | AW |
Remote access to clinical applications (IMAGE PROCESSING) | Optional |
Dicom 3-d image export (IMAGE PROCESSING) | Yes |
RECOMMENDED ROOM SIZE, m² (IMAGE PROCESSING) | 45.5 minimum |
RADIATION DOSE | |
Dose modulation technique (X-ray tube anode) | Volara DAS, OptiDose, color coding for kids, 3-D dose modulation, ECG dose modulation, beam tracking, short gantry geometry |
Pediatric specific dose control (RADIATION DOSE) | Color coding for kids |
Prospective ECG gating (RADIATION DOSE) | Optional |
Retrospective ECG editing (RADIATION DOSE) | Optional |
Axial cardiac (RADIATION DOSE) | Optional |
Low dose cardiac (axial acquisition) (RADIATION DOSE) | Optional |
Arrhythmia correction (RADIATION DOSE) | Optional |
IMAGE RECONSTRUCTION | |
Cpu (NETWORK/DATABASE) | Open architecture/ LINUX |
Scan fovs, cm (IMAGE PROCESSING) | 25, 50 |
Reconstruction matrices (IMAGE RECONSTRUCTION) | 512 x 512 |
Per slice, sec (Reconstruction time) | Up to 16 frames/sec |
No. of online images (IMAGE STORAGE) | 250,000 (512 x 512) |
Archive (IMAGE STORAGE) | 2.3 GB MOD, DICOM 3.0 |
kW output (GENERATOR) | 100 |
Kvp range (GENERATOR) | 80, 100, 120, 140 |
CLINICAL APPLICATIONS AND FUNCTIONALITY | |
Coronary artery calcification scoring (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Optional |
Quantification (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Yes |
Lung nodule assisted reading (Myocardial evaluation) | Optional, Lung VCAR |
Respiratory gating (DISPLAY) | Optional |
Virtual colonoscopy assisted reading (Myocardial evaluation) | Optional |
Vessel analysis (noncardiac) (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Optional, Vessel IQ Express |
Brain perfusion (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Optional |
Z-axis coverage for brain perfusion (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Optional, 80 mm Volume Shuttle |
Auto bone removal (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Optional |
Body perfusion (CLINICAL APPLICATIONS AND FUNCTIONALITY) | Optional |
Other (ROTORS) | Cardiac, perfusion, lung analysis, CT colonography, DentaScan, CT angiography |
DICOM (GANTRY) | As defined in DICOM Conformance Statement |
CT image storage SCU/SCP (Range of movement) | Yes |
Modality worklist SCU (Range of movement) | Yes |
Query/Retrieve SCU and SCP (Range of movement) | Yes |
Storage commitment SCU (Range of movement) | Yes |
Modality performed procedure step SCU (Range of movement) | Yes |
IHE profiles supported (Range of movement) | Yes |
DETECTOR (SCATTERED LIGHT) | |
Total detector width, z-axis, mm (DETECTOR) | 40 |
Reconstructed slice width options, mm (GANTRY) | 0.625, 1.25, 2.5, 3.75, 5, 7.5, 10 |
Standard Rotation times, sec, 360° (DETECTOR) | 0.35, 0.37, 0.4, 0.42, 0.45, 0.47, 0.5, 0.6, 0.7, 0.8, 0.9, 1, 2, 3, 4 |
OTHER ATTRIBUTES (Interference compensation) | AutoScan, AutoArchive, AutoFilm, AutoVoice, AutoTransfer, SmartPrep, ProtocolPro, View/Edit Wizard, DynaPlan Plus, ImageWorks, ProView, remote tilt, in-room start, rear gantry control, breathing lights, SmartmA-3-D dose modulation, ECG dose modulation wit |
MARKETING REGION (Interference compensation) | Worldwide |