| MODEL | HiSpeed CT/i |
| X-RAY TUBE | |
| X-RAY TUBE ANODE | |
| Heat storage, hu (X-ray tube anode) | 3500000 |
| Heat dissipation rate, hu/min (X-RAY TUBE) | 820000 |
| Tube cooling (X-ray tube anode) | Oil/air |
| Tube focal spot, mm (X-ray tube anode) | 0.4x 0.9, 0.6 x 0.9 |
| Optional x-ray tubes (X-ray tube anode) | Performix 630; 6,300,000HU |
| POWER NEEDED | 480VAC, 50/60 Hz, 3-phase delta |
| Type (CONFIGURATION) | Single slice helical |
| # of slices (X-ray tube anode) | 1 |
| DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC) | Yes |
| HELICAL SCANNING (DISPLAY) | Yes |
| Max scan time, sec (DISPLAY) | 60 |
| Max scan volume, cm (DISPLAY) | 170 |
| Spatial resolution, lp/cm (DISPLAY) | Same as axial |
| Pitch (DISPLAY) | 0.4:1 to 3:1 |
| Reconstruction time per image, sec (DISPLAY) | 3 (1.75optional) |
| Output, kw (X-RAY GENERATORS) | 48,optional 53 |
| Kvp range (GENERATOR) | 80.100.120.140 |
| Ma range (IMAGING SYSTEM) | 10440 with Performix |
| 3-D PACKAGE (DISPLAY) | Optional |
| FLOOR SPACE NEEDED | |
| Size, m2 (FLOOR SPACE NEEDED) | 28 (45optimal) |
| Image storage | |
| Hard disk, MB | 4.3GB |
| No. online images (IMAGE STORAGE) | 3,700 (512x 512), 450raw data |
| Archive (IMAGE STORAGE) | 2.3GB MOD, DICOM 3.0 |
| DISPLAY | |
| MONITOR | 20"(2) |
| Matrixes, pixels (DISPLAY) | 1024x 1024 |
| Range of ct numbers (DISPLAY) | -1,024to +3,071 |
| Image enlarging scale (DISPLAY) | Up to 8x |
| Max # slices displayed simultaneously (DISPLAY) | 16 |
| GENERATOR | |
| CT SIMULATION (DISPLAY) | Optional |
| IMAGE PROCESSING (MAXIMUM DISPLAY) | |
| Cpu (NETWORK/DATABASE) | Silicon GraphicsMIPS 4400/250 MHz |
| Scan fovs, cm (IMAGE PROCESSING) | 25.48 |
| Image reconstruction matrixes (IMAGE PROCESSING) | 256x 256, 512x 512 |
| Reconstruction time | |
| Per slice, sec (Reconstruction time) | 3,optional 1.8 |
| For localization scan, sec (Reconstruction time) | Real time |
| TABLE FEATURES | |
| Range of movement | |
| Vertical, cm (Range of movement) | 51-107 |
| Longitudinal, cm (Range of movement) | 170 |
| Scanning range,cm (Range of movement) | 170 |
| Max load capacity with accuracy, kg (Range of movement) | 180 (±0.25 mm), 205 (±1 mm) |
| PERFORMANCE | |
| Minimum interscan time, sec (PERFORMANCE) | 1 |
| Dynamic scan pace (PERFORMANCE) | 30scans/min |
| High-contrast spatial resolution 0%MTF, lp/cm (PERFORMANCE) | 15 |
| 50%MTF, lp/cm (PERFORMANCE) | 8.5 |
| Low-contrast resolution, mm at %at £4 rads (PERFORMANCE) | 2.5mm at 0.25% at 4rads AAPM w/ATS |
| Noise, %at £2.5 rads (PERFORMANCE) | 0.33@ 2.5 rads |
| CT ANGIOGRAPHY (DISPLAY) | Part of 3-D option |
| GANTRY | |
| Geometry (GANTRY) | Rotate-rotate, low-voltage slip ring |
| Number and type of detectors (GANTRY) | 840+ 12 reference, HiLight ceramic |
| # detection channels (GANTRY) | 744 |
| Rotation times, sec 360 (GANTRY) | 1.2.3.4; 0.8optional |
| Partial (GANTRY) | 0.5 |
| Slice thickness, mm (GANTRY) | 1,3,5,7,10 |
| X-ray fan beam angle, ° (GANTRY) | 49 |
| Gantry angle deg (GANTRY) | -30to +30 |
| Gantry size, hxwxd,cm (GANTRY) | 208x 223 x 109 |
| Gantry weight, kg (GANTRY) | 2979 |
| Gantry opening, cm (GANTRY) | 70 |
| Scan localizer (GANTRY) | Laser |
| OTHER ATTRIBUTES (Interference compensation) | Autoscan, autovoice, autofilm, autostore, autotransfer; Smart- Prep, ProtocolPro, view/edit wizard, DynaPlan Plus ImageWorks, ProView, PMR ConnectPro, Sherlock CBT, bar- code reader, Smart- Score, SmartView, Fast Recon. |
| FDA CLEARANCE (Interference compensation) | Yes |
| CE MARK (MDD) (Interference compensation) | Yes |
| MARKETING REGION (Interference compensation) | Worldwide |